FP40 / FP40-D Specifications
SINE SIGNAL
- Frequencies
Normal Sweep
- 1/12 octave frequencies from 200 to 8000 Hz, closest 100 Hz, within 1%
- Fast Sweep
- 1/3 octave frequencies closest 100 Hz
- Warbled Sinewave
- Has 5%, 33-1/3 Hz warble
- Short Sweep
- 1/2 octave frequencies closest 100 Hz
- Amplitude (RMS)
- Coupler mode: 40 dB SPL through 100 dB SPL in 5 dB steps. Probe mode: 40 dB SPL through 90 dB. Accuracy at reference point, after leveling, 2.5 dB for 500 Hz through 3500 Hz; 3.5 dB for all other frequencies.
- Harmonic Distortion
- (at 70 dB SPL) Less than 0.5% for 500, 800, & 1600 Hz.
COMPOSITE SIGNAL (optional)
- Frequencies
- From 200 Hz to 8000 Hz in 100 Hz intervals. Accuracy within 1%.
- Amplitude
- Coupler mode: (RMS) 40 dB SPL through 100 dB SPL in 5 dB steps.
-
- Probe mode: 40 dB SPL through 90 dB. Accuracy at reference point, after leveling, 2.5 dB for 0.5 kHz through 3.5 kHz; 3.5 dB for all other frequencies.
- Crest Factor
- Less than 9 dB.
BATTERY CURRENT MEASUREMENT
- Measurement Range
- 0 mA to 25.5 mA.
- Current Limit
- 55 mA.
- Accuracy
- 3% of full scale ±1 digit.
- Resolution
- 0.1 mA.
- Simulated Battery Types
- 10 A/230, 13, 312, 5, and 675 zinc air; 13, 312, 675, and 41 mercury; 13, 312, 76 silver; AA
- Battery Voltages Supplied
- 1.5 V for silver oxide and AA, 1.3 V for mercury and zinc air.
- Tolerance
- ±0.01V, no load.
DIGITAL READOUT OF SPL
- Frequency Range
- 200 Hz through 8000 Hz.
- Amplitude Range
- 0 dB SPL through 149.9 dB SPL, -70 dB through +100 dB gain.
- Max Input Signal
- 150 dB SPL.
- Resolution
- 0.1 dB
- Type
- True RMS.
- Accuracy
- From 250 Hz to 2500 Hz, 2 dB ±one digit. All other frequencies, 3 dB ±one digit.
HARMONIC DISTORTION ANALYSIS
- Type
- 2nd, 3rd, and 2nd + 3rd = total.
- Resolution .
- 1%
- Reading
- Percent with respect to total signal. Readings made at frequencies from 400 through 2500 Hz.
SYSTEM NOISE
- Equivalent Input Noise
- 50 dB SPL RMS.
- Noise Reduction
- Signal averaging synchronized with the signal generator. Averaging factors: off, 2, 4, 8, or 16. Random noise will be reduced by an amount equal to the inverse square root of the factor chosen.
POWER REQUIREMENTS
- Voltage
- 90 VAC to 264 VAC.
- Frequency
- 50 Hz to 60 Hz.
- Power Dissipation
- 40 VA at 120 VAC, 60 Hz input, normal operation, 55 VA while printing.
BATTERY OPERATION (optional)
- Operating Time:
- 3 hours continuous on battery power (with new battery at 25 degrees C).
- Auto Shutdown
- General shutdown after no operation of controls for 15 minutes (battery operation only).
- Battery Charger
- Built-in automatic battery charger. Full charge in 10 hours.
SOUND CHAMBER
- Test Area
- 3 1/2" x 5 1/2" (8.9 x 13.9 cm)
DISPLAY SCREEN
- Backlit Liquid Crystal Display
- Graphical display, 640 pixels wide x 200 pixels high. Display angle adjustable.
HIGH SPEED THERMAL PRINTER
- Print Speed
- Screen copy in 14-19 seconds.
- Paper Width
- 60 mm
EXTERNAL CONNECTORS
- Jacks
- RS232 (9-pin), probe monitor earphone (1/4" stereo), external speaker jack (3.5 mm stereo phone), VGA CRT (15 pin).
PHYSICAL DESCRIPTION
- Dimensions
- 20.125" x 14.750" x 6.5" (50.5 x 36.9 x 16.25 cm) (with lid on case).
- Color
- Light grey with black trim, white control panel.
- Weight
- 5 pounds (11,40 kg) with lid and battery; 22 pounds (10 kg) without battery.
GUARANTEE
The FONIX FP40 and its accessories are guaranteed to be free from manufacturing defects which would prevent the products from meeting these specifications for a period of one year from date of purchase.
Accessories and Options
FP40:
- Standard Accessories
- Operator's Manual
- HA-1, HA-2 couplers
- Battery pills #675/65, 312, 10A/230
- FM40 microphone
- Fun-Tak
- Ear level adapter
- Roll of printing tape
- Lid with accessories compartments
14mm to 1" adapter
- VGA monitor capability
-
- Standard with Probe Option
- Monitor headset
- Wedge ear hook
- Calibration clip
- Probe calibration adapter
- M250 probe microphone assembly
- Set of 25 probe tubes
- Felt pen-dry erase
- Mounting sleeve
-
- Optional Accessories
- External printer kit
- Swing arm, speaker and cable
Telewand for ANSI 96
Telecoil induction board
- Quest CQ-10 sound level calibrator
- Maintenance Manual (on request at time of purchase)
- Children's wedge ear hook
- 6010 sound chamber
- Battery pills #41, 5/5A, AA
- FM kit
VGA monitor
- Insert Earphone Kit
-
- Options
- Composite Option
- Probe Option (Real Ear)
-
- Battery Pack Option
- ID Option
- RS232 Option
- OES (Occluded Ear Simulator) Option
- CIC Option with coupler
- WinCHAP software
- IEC, ANSI 87, ANSI 96, JIS, ISI test sequences (one standard, others at extra cost.)
- DSL/Link programming
- Safety Approval UL 544 available on request at no additional charge
FP40-D:
- Standard Accessories
(Same as FP40, with these exceptions:)
-
- Add:
- All accessories standard with FP40 Probe Option except monitor headset.
- Exclude:
- Battery pills
Lid with accessories compartments
- Carrying handle
-
- Optional Accessories
(Same as FP40, with these exceptions:)
- Add:
- Battery pills #13, 675/76, 312, 10/230
- Soft carrying case
-
-
- Options
(Same as FP40, with these exceptions:)
Exclude:
- Probe Option (standard on FP40-D)
- Battery Pack
|
|